Health product recall

IRISpec CA/CB/CC

Last updated

Summary

Product
IRISpec CA/CB/CC
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

IRISpec CA/CB/CC

432-24 and higher

800-7702

Issue

Beckman Coulter has become aware of an increase in complaints for intermittent IRISpec CB failures for ketone on the iChemVELOCITY Urine Chemistry System (instructions for use PN 300-4449FH) and DxU 810c IRIS Urine Chemistry System (instructions for use PN C49355AA).
If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples. 

Recall start date: July 16, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter Ireland, Inc.

Lismeehan, Co. Clare, Ireland

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75875

Get notified

Receive emails about new and updated recall and safety alerts.