IRISpec CA/CB/CC
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
IRISpec CA/CB/CC |
432-24 and higher |
800-7702 |
Issue
Beckman Coulter has become aware of an increase in complaints for intermittent IRISpec CB failures for ketone on the iChemVELOCITY Urine Chemistry System (instructions for use PN 300-4449FH) and DxU 810c IRIS Urine Chemistry System (instructions for use PN C49355AA).
If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples.
Recall start date: July 16, 2024
Additional information
Details
Beckman Coulter Ireland, Inc.
Lismeehan, Co. Clare, Ireland