Health product recall

iRICELL, IQ 200 and DxU Urine Analyzers

Last updated

Summary

Product
iRICELL, IQ 200 and DxU Urine Analyzers
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
iRICELL 2000 Plus 7871 800-7714
iRICELL 3000 Plus 7871 800-3079 800-7713
IQ 200 Sprint Plus W Load/Unload 7871 800-3081
iRICELL 2000 Pro 7871 800-3049
DxU 850m/840m Microscopy Analyzers 100188 C76947
IQ 200 Sprint Pro 7871 800-3053
IQ 200 Sprint Pro W Load/Unload 7871 800-3082
IQ 200 Sprint Plus 7871 800-3933
iRICELL3000 Pro 7871 800-3080 800-3050

Issue

Beckman Coulter has become aware that the cannula, which is found in the flowcell and the IQclear rinse waste well assembly, may rust. Rust from the cannula has the potential to generate erroneous results.

Recall start date: December 18, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter Ireland, Inc.

Lismeehan, Co. Clare, Ireland

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75031

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe