Product
iRICELL, IQ 200 and DxU Urine Analyzers
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
iRICELL 2000 Plus | 7871 | 800-7714 |
iRICELL 3000 Plus | 7871 | 800-3079 800-7713 |
IQ 200 Sprint Plus W Load/Unload | 7871 | 800-3081 |
iRICELL 2000 Pro | 7871 | 800-3049 |
DxU 850m/840m Microscopy Analyzers | 100188 | C76947 |
IQ 200 Sprint Pro | 7871 | 800-3053 |
IQ 200 Sprint Pro W Load/Unload | 7871 | 800-3082 |
IQ 200 Sprint Plus | 7871 | 800-3933 |
iRICELL3000 Pro | 7871 | 800-3080 800-3050 |
Issue
Beckman Coulter has become aware that the cannula, which is found in the flowcell and the IQclear rinse waste well assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
Recall start date: December 18, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Beckman Coulter Ireland, Inc.
Lismeehan, Co. Clare, Ireland
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75031