iRICELL, IQ 200 and DxU Urine Analyzers
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
iRICELL 2000 Plus |
7871 |
800-7714 |
iRICELL 3000 Plus |
7871 |
800-3079 |
IQ 200 Sprint Plus W Load/Unload |
7871 |
800-3081 |
iRICELL 2000 Pro |
7871 |
800-3049 |
DxU 850m/840m Microscopy Analyzers |
100188 |
C76947 |
IQ 200 Sprint Pro |
7871 |
800-3053 |
IQ 200 Sprint Pro W Load/Unload |
7871 |
800-3082 |
IQ 200 Sprint Plus |
7871 |
800-3933 |
iRICELL3000 Pro |
7871 |
800-3080 |
Issue
Beckman Coulter has become aware that the cannula, which is found in the flowcell and the IQclear rinse waste well assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
Recall start date: December 18, 2023
Additional information
Details
Beckman Coulter Ireland, Inc.
Lismeehan, Co. Clare, Ireland
Get notified
Receive notifications for new and updated recalls and alerts by category.