Health product recall

Iq Calibrator Pack and Control/Focus Set

Last updated

Summary

Product
Iq Calibrator Pack and Control/Focus Set
Issue
Medical devices - Quality issue
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products Lot or Serial Number Model or catalog number
Iq Control/Focus Set 146-21 800-3104
Iq Calibrator Pack 146-21 800-3103

Issue

Beckman Coulter has identified that order number(s) 41231574 and 41231557 for the shipment of Iq200 Control/Focus Set (part number 800-3104, lot# 146-21) and Iq200 Calibrator (part number 800-3103, lot# 146-21) were inadvertently frozen prior to shipment. There is a potential that the red blood cells (RBC) in the products will degrade after a freeze/thaw cycle.

Recall start date: Nov 8, 2021

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Beckman Coulter, Inc.

250 S. Kraemer Blvd., Brea, California

United States, 92821

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63980

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe