IOMAX Cortical Module
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Return affected devices to manufacturer for repair or replacement.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
IOMAX Cortical Module |
19029603AA0124001 |
190296-200 |
Issue
An assembly error was discovered on the cortical module (PN 190296-200). A cortical module insulator that is used to provide an electric barrier between circuit boards for TCS-9, ICS, and other internal parts was found to be installed incorrectly. The insulator provides clearance between safety critical components and incorrect installation does not meet the minimum creepage/air gap required to meet IEC 60601-1.
Recall start date: April 23, 2024
Additional information
Details
Cadwell Industries, Inc.
909 North Kellogg Street, Kennewick, Washington, United States, 99336