Health product recall

Invos Reusable Sensor Cable For PM7100

Brand(s)
Last updated

Summary

Product
Invos Reusable Sensor Cable For PM7100
Issue
Medical devices - Performance
What to do

Contact the importer GE Healthcare Technologies Canada. 

Audience
Healthcare
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
Invos Reusable Sensor Cable For PM7100 2023-04-28 2023-05-04 PMAC71RSC

Issue

It was discovered that some reusable sensor cables contain manufacturing defects which can result in the "check sensor" or "replace sensor" error messages which result in a non-functioning channel that cannot obtain a regional oxygen saturation (rSo2) reading when using an affected cable.

Recall start date: May 3, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Covidien Llc
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-75573

Get notified

Receive notifications for new and updated recalls and alerts by category.