This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Invacare Carroll CS7 Bed (2014-12-19)

Starting date:
December 19, 2014
Posting date:
January 7, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43147

Recalled Products

Invacare Carroll CS7 Bed

Reason

A potential defect involving the Invacare Carroll CS7 bed, model IHCS7HLOCK-C. The potential defect will not allow assembly of 36"" wide CS assist rails, 36"" wide CS assist bars & 68"" partial length rail onto the bed. The mounting bracket on the bed was incorrectly welded by 180 degrees which does not allow for the rail accessories to be installed onto the bed. The defective mounting rail brackets are located on the frame of the bed on both sides. The bottom square hole of the bracket does not align with the rail or rail accessories.

Affected products

Invacare Carroll CS7 Bed

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

IHCS7HLOCK-C

Companies
Manufacturer

Invacare Corp.

One Invacare Way

Elyria

44036

Ohio

UNITED STATES