Intouch Bed Caster Stem (2014-07-02)

Starting date:
July 2, 2014
Posting date:
July 22, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40565

Recalled Products

Intouch Bed Caster Stem 

Reason

Stryker received complaints which allege that a caster has broken on the top of the caster stem where the brake rod inserts into the caster. If more than one caster on a unit is damaged, brake functions may be affected and result in reduced brake holding force. If only one caster on the bed is damaged, brake functionality continues to meet specifications.

Affected products

Intouch Bed Caster Stem

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number
  • 2130
  • 2131
  • 2140
  • 2141
Companies
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
Michigan
UNITED STATES