Intouch Bed Caster Stem (2014-07-02)
- Starting date:
- July 2, 2014
- Posting date:
- July 22, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-40565
Recalled Products
Intouch Bed Caster Stem
Reason
Stryker received complaints which allege that a caster has broken on the top of the caster stem where the brake rod inserts into the caster. If more than one caster on a unit is damaged, brake functions may be affected and result in reduced brake holding force. If only one caster on the bed is damaged, brake functionality continues to meet specifications.
Affected products
Intouch Bed Caster Stem
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 2130
- 2131
- 2140
- 2141
Companies
- Manufacturer
-
Stryker Medical
3800 E Centre Avenue
Portage
49002
Michigan
UNITED STATES