Intellivue Information centre IX - software (2015-04-13)
- Starting date:
- April 13, 2015
- Posting date:
- May 8, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-53277
Recalled Products
A. Intellivue Information centre IX - software
Reason
Philips has become aware that when HL7 lab messages are sent to a PIIC iX system that has been localized for comma delimiters, the lab result is incorrectly displayed on the bedside monitor. An expected delimiter is stripped from the lab value and the number shown is of a higher magnitude than expected. If lab results are incorrect, a clinician may initiate incorrect treatment or delay necessary treatment this problem only happens when the PIIC iX is configured during set up for regional settings where the delimiter is a comma. The lab option must be in use, and the workflow utilizes the automatic lab interface displayed at the bedside monitor.
Affected products
A. Intellivue Information centre IX - software
Lot or serial number
1M4H-34MB-4
4744-35J4-F
577U-56K0-9
5G4L-1CJJ-0
7U00-2CJW-V
7Z4V-37N1-L
Model or catalog number
M3290B
Companies
- Manufacturer
-
Philips Medical Systems
3000 Minuteman Road
Andover
Massachusetts
UNITED STATES