This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

IntelliCuff

Starting date:
June 7, 2017
Posting date:
July 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63846

Reason

IntelliCuff standalone device motor may cease to function during use causing alarm sounds and red LEDS to blink in all segments. There is no risk for the patient or user related to this failure.

Affected products

IntelliCuff

Lot or serial number

Lots from July 2015 to present

Serial numbers between 1000 - 2090

Model or catalog number

Not Applicable

Companies

Manufacturer

Hamilton Medical AG

Via Crush 8

Bonaduz

7402