Health product recall

Inteliq Specialty Immunoassay Control

Last updated

Summary

Product
Inteliq Specialty Immunoassay Control
Issue
Medical devices
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Inteliq Specialty Immunoassay Control 64954T 64952T 12008278 12008276

Issue

Bio-Rad laboratories would like to inform you that our Inteliq specialty immunoassay control (catalogue number 12008276, lot 64954t) and (catalogue number 12008278, lot 64954t) shows a stability issue. We have observed that PTH (intact) may not meet the thawed unopened, refrigerator off-board and instrument on-board storage stability claims as printed in the package insert.

Recall start date: May 24, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Bio-Rad Laboratories

9500 Jeronimo Road, Irvine, California, United States, 92618

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64236

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe