Health product recall

Intelepacs - Main Unit (2014-06-13)

Starting date:
June 13, 2014
Posting date:
July 3, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40275

Recalled Products

Intelepacs - Main Unit

Reason

In specific situations a partially truncated file may be written to  the Intelepacs server. This partially truncated file can still be viewed by inteleviewer and inteleviewer will report to the user that the file in question is complete, when in fact they were looking at a glossy image.

Affected products

Intelepacs - Main Unit

Lot or serial number

Versions 3-8-1 to 4-9-1

Model or catalog number

Intelepacs V3-X-X

Companies
Manufacturer

Intelerad Medical Systems Incorporated

895 De La Gauchetiere Street West, 4th floor

Montreal

H3B 4G1

Quebec

CANADA