Health product recall

IntelePACS (Image Fusion Module)

Last updated

Summary

Product
IntelePACS (Image Fusion Module)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
IntelePACS (Image Fusion Module) Software version 5-6-1-P23 and later INTELEPACS

Issue

An incorrect SUV (standard uptake value) calculation occurs in the PET/CT fusion viewer module. This issue only occurs on days with a DST (daylight savings time) change. The issue occurs because the calculation relies on a time calculation which does not correctly consider the DST change. An incorrect SUV may lead to an incorrect diagnosis.

Recall start date: February 4, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Intelerad Medical Systems Incorporated

800 Boulevard De Maisonneuve Est, Montreal, Quebec, Canada, H2L 4L8

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76978

Get notified

Receive notifications for new and updated recalls and alerts by category.