IntelePACS (Image Fusion Module)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
IntelePACS (Image Fusion Module) |
Software version 5-6-1-P23 and later |
INTELEPACS |
Issue
An incorrect SUV (standard uptake value) calculation occurs in the PET/CT fusion viewer module. This issue only occurs on days with a DST (daylight savings time) change. The issue occurs because the calculation relies on a time calculation which does not correctly consider the DST change. An incorrect SUV may lead to an incorrect diagnosis.
Recall start date: February 4, 2025