IntelePACS
Summary
Contact the manufacturer if you need additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
|---|---|---|
IntelePACS |
EV.1.5.5 to EV.1.19.14 |
INTELEPACS |
Issue
|
The issue may be triggered by the presence of pre-formatted html text in HL7 reports using the <pre> tag. These reports will be truncated when they are rendered by the enhanced viewer pdf report viewer, which could lead to incomplete report content displayed to the enhanced viewer user. This issue will only happen while using Microsoft Edge/Chromium 92, Mozilla Firefox 90, Google Chrome 92 and later versions of these browsers. Recall start date: 2021-10-29 |
What you should do
Contact the manufacturer if you need additional information.
Additional information
Details
|
INTELERAD MEDICAL SYSTEMS INCORPORATED 800 BOULEVARD DE MAISONNEUVE EST, 12IEME ETAGE, H2L 4L8
|
Get notified
Receive emails about new and updated recall and safety alerts.