Health product recall

IntelePACS

Last updated

Summary

Product
IntelePACS
Issue
Medical devices - Software defect
What to do

Contact the manufacturer if you need additional information.

Affected products

Affected Products Lot or serial number Model or catalog number
IntelePACS EV.1.5.5 to EV.1.19.14 INTELEPACS

Issue

The issue may be triggered by the presence of pre-formatted html text in HL7 reports using the <pre> tag. These reports will be truncated when they are rendered by the enhanced viewer pdf report viewer, which could lead to incomplete report content displayed to the enhanced viewer user. This issue will only happen while using Microsoft Edge/Chromium 92, Mozilla Firefox 90, Google Chrome 92 and later versions of these browsers.

Recall start date: 2021-10-29

What you should do

Contact the manufacturer if you need additional information.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

INTELERAD MEDICAL SYSTEMS INCORPORATED

800 BOULEVARD DE MAISONNEUVE EST, 12IEME ETAGE, H2L 4L8

MONTREAL, QUEBEC  CANADA
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63598

Get notified

Receive notifications for new and updated recalls and alerts by category.