Instrument Manager (2021-09-04)
- Starting date:
- September 4, 2021
- Posting date:
- September 21, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76467
Last updated:
2021-09-21
Affected Products
Instrument Manager
Reason
After the upgrade of the data innovations advanced configurable delimited instrument driver to V8.04.0032, It was observed that the driver is not reading timestamps correctly when reading and parsing date/time values from a results file. When there are multifield date/time values for results, the driver is parsing both values but is discarding the time value resulting in a timestamp of 00:00:00 (midnight).
Affected products
Instrument Manager
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
V8.04 AND HIGHER
Companies
- Manufacturer
-
Data Innovations LLC
120 Kimball Avenue Suite 100
South Burlington
05403
Vermont
UNITED STATES