Health product recall

Instinct® Plus Endoscopic Clipping Device

Last updated

Summary

Product
Instinct® Plus Endoscopic Clipping Device
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalogue number

Instinct® Plus Endoscopic Clipping Device

More than 10 numbers, contact manufacturer.

INSC-P-7-230-S

Issue

Cook Medical has received 486 complaints of a malfunction that occurs when the user attempts to open the clip jaws by actuating the handle. The malfunction that may occur is the clip housing detaching from the catheter attachment and the clip remaining attached to the drive wire. The clip will extend from the catheter, but it will remain attached to the internal drive wire instead of opening the clip jaws. The clip cannot be opened if this occurs. 

Corrective actions have been implemented to mitigate the risk of this failure mode.  This action is being taken to withdraw products manufactured prior to the implementation of the corrective actions.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Wilson-Cook Medical, Inc.

4900 Bethania Station Rd., Winston Salem, North Carolina, United States, 27105

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-81736