INOmax DS
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
INOmax DS |
More than 10 numbers, contact manufacturer. |
10007 |
Issue
The INOmax DS IR delivery system, is not compatible with the upgraded version of the Leoni plus ventilator, manufactured by Lowenstein Medical. the Leoni plus ventilator was recently upgraded by Lowenstein Medical, with new hardware and software, and the upgraded version cannot be appropriately connected with the INOmax DS IR as instructed in the INOmax DS IR operators manual (om). the INOmax DS IR om requires warns that the one-way valve must be used.
Recall start date: June 29, 2022
Additional information
Details
Mallinckrodt Manufacturing Llc
6603 Femrite Drive, Madison, Wisconsin, United States, 53718