This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Innova E4 and E5 CMS (2017-06-14)

Starting date:
June 14, 2017
Posting date:
July 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63882

Affected Products

A.  Innova E4 CMS

B. Innova E5 CMS

Reason

This recall is being initiated by the manufacturer to reduce the chance of a blockage and improve the flow of water and chemical through the nozzle within the unit that would connect the 7th channel in a 7-channel scope to the Innova unit. The size of this nozzle is currently 0.2 mm and it will be replaced by a 0.5 mm nozzle.

Affected products

A.  Innova E4 CMS

Lot or serial number

59742001

59742002

59742003

59742004

60361001

60361002

60361003

Model or catalog number

50600.92

50600.93

E4-CMS

E4CMS-DH

Companies
Manufacturer

BHT HYGIENETECHNIK GMBH

MESSERSCHMITTSTRASSE 11

GERSTHOFEN

86368

GERMANY



B. Innova E5 CMS

Lot or serial number

68881001

68881002

68881003

68881004

69880001

69880002

69880003

Model or catalog number

50600.94

50600.95

50600.96

50600.97

E5-CMS

E5CMS-DC-EFF

E5CMS-DH

E5CMS-DH-EFF

Companies
Manufacturer

BHT HYGIENETECHNIK GMBH

MESSERSCHMITTSTRASSE 11

GERSTHOFEN

86368

GERMANY