This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

INGENUITY CT SYSTEM (2016-04-26)

Starting date:
April 26, 2016
Posting date:
May 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58424

Affected Products

INGENUITY CT SYSTEM

Reason

Two different length fasteners were found to be made of stainless steel rather than the specified alloy steel. The nonconforming fasteners continue to meet the tensile safety factor requirements for the application. However given the material characteristics of stainless steel, these fasteners may not be reused (re-torqued).

Affected products

INGENUITY CT SYSTEM

Lot or serial number

Not applicable

Model or catalog number
  • 728323
Companies
Manufacturer

Philips Medical Systems (Cleveland) Inc.

595 MINER ROAD

CLEVELAND

Ohio

UNITED STATES