Ingenuity CT System (2014-03-26)

Starting date:
March 26, 2014
Posting date:
April 15, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39095

Recalled products

  1. Ingenuity CT System

Reason

Philips has become aware of the following two (2) potential issues that may result in the subframe of the table top becoming free floating and causing unintended, horizontal motion. The two potential issues are:

  1. Service latch may not be secured.
  2. Service latch threaded rod may be fractured.

Affected products

A. Ingenuity CT System

Lot or serial number
  • 320148
Model or catalog number
  • 728323