This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Ingenuity CT System (2014-03-26)

Starting date:
March 26, 2014
Posting date:
April 15, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39095

Recalled products

  1. Ingenuity CT System

Reason

Philips has become aware of the following two (2) potential issues that may result in the subframe of the table top becoming free floating and causing unintended, horizontal motion. The two potential issues are:

  1. Service latch may not be secured.
  2. Service latch threaded rod may be fractured.

Affected products

A. Ingenuity CT System

Lot or serial number
  • 320148
Model or catalog number
  • 728323
Companies
Manufacturer

Philips Medical Systems (Cleveland), Inc.

595 Miner Road

Cleveland

44143

Ohio

UNITED STATES