Ingenuity CT System (2014-03-26)
- Starting date:
- March 26, 2014
- Posting date:
- April 15, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39095
Recalled products
- Ingenuity CT System
Reason
Philips has become aware of the following two (2) potential issues that may result in the subframe of the table top becoming free floating and causing unintended, horizontal motion. The two potential issues are:
- Service latch may not be secured.
- Service latch threaded rod may be fractured.
Affected products
A. Ingenuity CT System
Lot or serial number
- 320148
Model or catalog number
- 728323
Companies
- Manufacturer
-
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES