Health product recall

Ingenia Elition X and S (2018-12-06)

Starting date:
December 6, 2018
Posting date:
January 8, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68688

Affected products

  1. Ingenia Elition X
  2. Ingenia Elition S

Reason

Philips has become aware of a potential hazardous situation with the VitalScreen 'Stop Table' software button on the affected products (the yellow button on the right bottom corner). During automatic tabletop movement, pressing and holding the 'Stop Table' button may lead to an unexpected restart of the table movement.

Affected products

A. Ingenia Elition X

Lot or serial number

Not applicable

Model or catalog number
  • 781358
Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS



A. Ingenia Elition X

Lot or serial number

Not applicable

Model or catalog number
  • 781357
Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS