This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Infinix CF-1/BP-System (2014-04-04)

Starting date:
April 4, 2014
Posting date:
April 30, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39223

Recalled products

  1. Infinix CF-1/BP-System

Reason

It has been found that even if the touch sensor for the flat panel detector operates normally during the movement of tabletop longitudinal tilt, the movement does not stop.

Affected products

A. Infinix CF-1/BP-System

Lot or serial number

Serial Number:

  • MDB1052001
Model or catalog number
  • INFX-8000V / MD
Companies
Manufacturer

Toshiba Medical Systems Corp.

1385 Shimoishigami, Otawara-Shi

Tochigi-Ken

324-8550

JAPAN