Infinity® Centralstation Wide
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Infinity® Centralstation Wide |
More than 10 numbers, contact manufacturer. |
MS26800 |
Issue
The recall is being initiated because the new edifier speaker configured with the ICS in analog mode, may not work as expected after a power loss or power (usb cable) being disconnected. The new edifier speaker configured with ICS in analog mode will default back to usb mode, after power being restored or power cable (usb cable) reconnected which can lead to the loss of audio alarms.
Recall start date: December 23, 2024
Additional information
Details
Draeger Medical Systems, Inc.
3135 Quarry Road, Telford, Pennsylvania, United States, 18969
Get notified
Receive emails about new and updated recall and safety alerts.