Health product recall

Infinity Occipitocervical Upper Thoracic System Multi Axial Screw

Last updated

Summary

Product
Infinity Occipitocervical Upper Thoracic System Multi Axial Screw
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Infinity Occipitocervical Upper Thoracic System Multi Axial Screw H5802983 3604024

Issue

It was identified by Medtronic that an incorrect type of titanium was used for the manufacturing of subcomponents of multi-axial screws that are for use as part of the Infinity oct system. This incorrect material reduces the clamping strength of the multi-axial screw, which may lead to a loss of fixation with the infinity system construct.

Recall start date: February 16, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Medtronic Sofamor Danek Usa, Inc.

1800 Pyramid Place, Memphis, Tennessee, United States, 38132

Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-73005

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe