Health product recall

Infinity M540 (2018-03-19)

Starting date:
March 19, 2018
Posting date:
April 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66412

Affected products

Infinity M540

Reason

Draeger learned about an issue with the infinity 540 patient monitor with software versions vg4.1 and below. When the m540 patient monitor is used with and connected to the Infinity MCable Mainstream CO2 (m11.1) sensor revision 16, the m540 patient monitor will reboot if the sensor enters a reduced accuracy mode. If the m540 patient monitor reboots 3 times within a 10 minute period or less, the m540 patient monitor will go enter a fail-state, resulting in a resetting to factory defaults and loss of patient settings and stored patient data. Revision 16 sensors were introduced in mid-2017. Use of revision 15 and earlier sensors will not cause this issue to occur.

Affected products

Infinity M540

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number
  • MS20401
  • MS22956
  • MS25643
  • MS26196
  • MS31818
Companies
Manufacturer
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
UNITED STATES