Infinity M540 (2018-03-19)
- Starting date:
- March 19, 2018
- Posting date:
- April 13, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66412
Affected products
Infinity M540
Reason
Draeger learned about an issue with the infinity 540 patient monitor with software versions vg4.1 and below. When the m540 patient monitor is used with and connected to the Infinity MCable Mainstream CO2 (m11.1) sensor revision 16, the m540 patient monitor will reboot if the sensor enters a reduced accuracy mode. If the m540 patient monitor reboots 3 times within a 10 minute period or less, the m540 patient monitor will go enter a fail-state, resulting in a resetting to factory defaults and loss of patient settings and stored patient data. Revision 16 sensors were introduced in mid-2017. Use of revision 15 and earlier sensors will not cause this issue to occur.
Affected products
Infinity M540
Lot or serial number
More than 100 numbers, contact manufacturer.
Model or catalog number
- MS20401
- MS22956
- MS25643
- MS26196
- MS31818
Companies
- Manufacturer
-
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
UNITED STATES