Health product recall

INFINITY ACUTE CARE WORKSTATION (2021-08-17)



Starting date:
August 17, 2021
Posting date:
September 10, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76407


Last updated:

Affected Products

  1. INFINITY ACUTE CARE WORKSTATION CRITICAL CARE
  2. INFINITY ACUTE CARE SYSTEM WORKSTATION NEONATAL CARE

Reason

Draeger has become aware of Sporadic cases in which the affected products with sw2.60 And lower with installed co2 measurement option restarted.

A restart is a designed system recovery function to restore the proper function of the ventilator. In the rare reported cases, however, an error in the data processing of the activated Co2 measurement caused the restart of the ventilation unit. Devices without integrated or with disabled co2 measurement function are not affected. The behavior can be effectively prevented by not activating the integrated Co2 measurement. If necessary, external Co2 monitoring can be used.

Affected products

  1. INFINITY ACUTE CARE WORKSTATION CRITICAL CARE

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

8416400

Companies
Manufacturer

Draegerwerk AG & Co. KGaA

Moislinger Allee 53-55

Lübeck

23542

GERMANY



  1. INFINITY ACUTE CARE SYSTEM WORKSTATION NEONATAL CARE

Lot or serial number
  • ASHN-0067
  • ASNN-0115
Model or catalog number

8417400

Companies
Manufacturer

Draegerwerk AG & Co. KGaA

Moislinger Allee 53-55

Lübeck

23542

GERMANY