INFINITY ACUTE CARE WORKSTATION (2021-08-17)
- Starting date:
- August 17, 2021
- Posting date:
- September 10, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76407
Last updated:
2021-09-10
Affected Products
- INFINITY ACUTE CARE WORKSTATION CRITICAL CARE
- INFINITY ACUTE CARE SYSTEM WORKSTATION NEONATAL CARE
Reason
Draeger has become aware of Sporadic cases in which the affected products with sw2.60 And lower with installed co2 measurement option restarted.
A restart is a designed system recovery function to restore the proper function of the ventilator. In the rare reported cases, however, an error in the data processing of the activated Co2 measurement caused the restart of the ventilation unit. Devices without integrated or with disabled co2 measurement function are not affected. The behavior can be effectively prevented by not activating the integrated Co2 measurement. If necessary, external Co2 monitoring can be used.
Affected products
-
INFINITY ACUTE CARE WORKSTATION CRITICAL CARE
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
8416400
Companies
- Manufacturer
-
Draegerwerk AG & Co. KGaA
Moislinger Allee 53-55
Lübeck
23542
GERMANY
-
INFINITY ACUTE CARE SYSTEM WORKSTATION NEONATAL CARE
Lot or serial number
- ASHN-0067
- ASNN-0115
Model or catalog number
8417400
Companies
- Manufacturer
-
Draegerwerk AG & Co. KGaA
Moislinger Allee 53-55
Lübeck
23542
GERMANY