Health product recall

Infinity Acute Care System Workstation (2014-02-24)

Starting date:
February 24, 2014
Posting date:
March 25, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38581

Recalled Products

Infinity Acute Care System Workstation

Reason

The battery capacity of the optional PS500 of the Infinity ACS workstation critical care (used in connection with Evita Infinity V500, Babylog VN500 or Evita V300) will not last as expected. In some cases neither the "battery low" nor the "battery depleted" alarm was given when the remaining battery capacity fell below 10%. However, when the battery depleted toatlly, the power fail alarm was generated.

Affected products

Infinity Acute Care System Workstation

Lot or serial number

8417400

Model or catalog number

More than 10 numbers, contact manufacturer

Companies
Manufacturer
Draeger Medical GMBH
Moislinger Allee 53-55,
Lubeck
23542
GERMANY