Infinity Acute Care System Workstation (2014-02-24)
- Starting date:
- February 24, 2014
- Posting date:
- March 25, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-38581
Recalled Products
Infinity Acute Care System Workstation
Reason
The battery capacity of the optional PS500 of the Infinity ACS workstation critical care (used in connection with Evita Infinity V500, Babylog VN500 or Evita V300) will not last as expected. In some cases neither the "battery low" nor the "battery depleted" alarm was given when the remaining battery capacity fell below 10%. However, when the battery depleted toatlly, the power fail alarm was generated.
Affected products
Infinity Acute Care System Workstation
Lot or serial number
8417400
Model or catalog number
More than 10 numbers, contact manufacturer
Companies
- Manufacturer
-
Draeger Medical GMBH
Moislinger Allee 53-55,
Lubeck
23542
GERMANY