Health product recall

Infant/Neonatal and Adult/Pediatric Filterline and Vitaline H sets, Airway Adapters, Ventline Set, and CO2 Sampling Line

Last updated

Summary

Product
Infant/Neonatal and Adult/Pediatric Filterline and Vitaline H sets, Airway Adapters, Ventline Set, and CO2 Sampling Line
Issue
Medical devices - Labelling and packaging
What to do

​​​​​​Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Filterline Set All lots. XS04620
Filterline H Set Infant/Neonatal All lots. 006324
Omni Ventline Set All lots. 012495
Filterline H Set All lots. XS04624
Filterline H Set Adult/Pediatric All lots. 007737 010580
Airway Adapter, Adult/Pediatric All lots. 010989
CO2 Sampling Line All lots. 010991
Filterline H Set Infant/Neonatal, Long All lots. 007738
Filterline Set Long, Adult/Pediatric All lots. 010579 007768
Vitaline H Set Adult/Pediatric All lots. 010787
Vitaline H Set Infant/Neonatal All lots. 010807

Issue

Medtronic is issuing an update to Instruction For Use (IFU) following reports of difficulty or inability in disconnecting the adapter from patient's endotracheal tube to perform a procedure.  As noted in the IFU, during setup, ensure that the airway adapter can be easily attached and detached from the breathing circuit/tubing before proceeding.

Recall Start Date: February 19, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Oridion Medical 1987 Ltd.

7 Hamarpe St., Jerusalem, Israel, 9777407

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77070