This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Infant Warmer system (2016-08-09)

Starting date:
August 9, 2016
Posting date:
August 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59898

Affected products

  1. Infant Warmer system
  2. Infant Warmer System (IWS)

Reason

Hot screws from the heater head of the infant warmer system could fall onto the bed if the "heater head" assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

Affected products

  1. Infant Warmer system

Lot or serial number

Not Applicable

Model or catalog number

3051

Companies
Manufacturer

Ohmeda Medical, A Division Of Datex Ohmeda Inc.



B. Infant Warmer System (IWS)

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

2001

3000

3050

3100

3300

3400

3500

4000

4300

4400

5000

Companies
Manufacturer

Ohmeda Medical, A Division Of Datex Ohmeda Inc.