This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Infant Radiant Warmers (2014-11-18)

Starting date:
November 18, 2014
Posting date:
November 28, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42685

Recalled Products

Infant Radiant Warmers

Reason

Fisher & Paykel Healthcare is issuing this voluntary recall notice because it has had two confirmed reports of breakage of the nut that secures the heater head in place. Breakage of the nut can cause the heater head to become partially detached and swing towards the bassinet. The heater head did not contact the patient in either case. However, there is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.

Affected products

Infant Radiant Warmers

Lot or serial number

071116 to 090810
071208002349 to 090717000930

Model or catalog number

More than 10 numbers, contact manufacturer.

Companies
Manufacturer
Fisher and Paykel Healthcare Ltd.
15 Maurice Paykel Place East Tamaki, P.O. Box 14348 Panmure
Auckland
2013
NEW ZEALAND