Inceptiv
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Inceptiv |
More than 10 numbers, contact manufacturer. |
977119 |
Issue
Performing a switch device operation from the 97725 or 9772501 WENS using the A71300 SCS trialing app to transfer data (e.g. patient and lead information) to a new implanted Inceptiv or Intellis Pro Implantable Neurostimulator (INS) results in the A71400 CP app crashing and the INS becoming unusable when the transfer involves a predefined legacy lead option selected in the app that uses the forward slash character "/".
Recall start date: May 28, 2025
Additional information
Details
| Medtronic Inc. |
| 710 Medtronic Parkway N.E., Minneapolis, Minnesota, United States, 55432 |
Get notified