Important Safety Information on the Interaction between Active Implantable Medical Devices and Systems and other Medical Devices - Notice to Hospitals
- Starting date:
- October 5, 2005
- Posting date:
- October 5, 2005
- Type of communication:
- Notice to Hospitals
- Subcategory:
- Medical Device
- Source of recall:
- Health Canada
- Audience:
- Hospitals
- Identification number:
- RA-170001085
Notice about Health Canada advisories
Notice to Hospitals - Health Canada Issued Important Safety Information on Active Implantable Medical Devices and Systems
An updated Notice to Hospitals has been posted on Health Canada website on December 20, 2005.
October 5, 2005
To: Hospital Chief of Medical Staff
Please distribute to the relevant Departments of Physiotherapy, Occupational Therapy, Radiology, Rehabilitative Medicine, Surgery, Neurosurgery, Orthopedic Surgery, Biomedical Engineering, Emergency Medicine, Pediatrics, Geriatrics, Internal Medicine, Nursing, Dentistry, Intensive Care and other involved professional staff and post this NOTICE in your institution.
Subject: Interactions between active implantable medical devices and systems and other medical devices
Patients with active implantable medical devices and systems (AIMDS) are at risk of injury from interactions between the AIMDS and other medical devices used for diagnostic or therapeutic purposes. Such interactions could damage some AIMDS resulting in loss of therapy from these units, and could cause tissue damage resulting in severe injury or death. Patients may be at risk even if the leads are not connected to an implantable device or even if the device is off.
On April 2, 2003, Heath Canada issued a Notice to Hospitals entitled "Important safety information regarding DIATHERMY THERAPY in patients with implanted leads and implanted systems with leads".Footnote 1 It warned about the risks of tissue overheating, serious injury and death in this group of patients when subject to shortwave (radiofrequency) or microwave diathermy therapy. Since then, Health Canada has become aware of further international reports of serious injury resulting from interactions between AIMDS and other medical devices.Footnote 2
In 2003, the French Health Products Safety Agency, Agence française de sécurité sanitaire des produits de santé (AFSSAPS), assembled a working group to analyse the recommendations of AIMDS manufacturers and make recommendations to health care professionals who perform certain diagnostic and therapeutic procedures on patients with AIMDS. The recommendations of the group were released by AFSSAPS in February 2005.Footnote 3
| Implanted Cardiac Pacemaker | Implanted Cardiac Defibrillator | Implanted Neurostimulator | |
|---|---|---|---|
|
Table 1 footnotes
|
|||
| Shortwave and Microwave Diathermy | Absolute contraindication | Absolute contraindication | Absolute contraindication |
| Ultrasound Diathermy | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Absolute contraindication |
| Electromagnetic Stimulation Device | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Absolute contraindication |
| MRI | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Absolute contraindication |
| Electrosurgery Device | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * |
| External Defibrillator | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * |
| Radiotherapy Device | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * |
| Lithotripsy or Ultrasound Therapy Device | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * | Take PrecautionTable 1 footnote * |
| Fluoroscopy or other X-ray Devices | No contraindication | No contraindication | No contraindication |
| Echography | No contraindication | No contraindication | No contraindication |
The risk of interactions between AIMDS and other medical devices used for diagnostic or therapeutic purposes depends on many parameters. Due to the diversity and complexity of the devices in question, Health Canada wishes to pass along these general guidelines but cannot give all-encompassing recommendations that cover all possible clinical scenarios. Health Canada continues to monitor this evolving safety issue and will release more information as it becomes available.
Reporting rates determined on the basis of spontaneously reported post-marketing adverse incidents are generally presumed to underestimate the risks associated with medical device use. Health Canada depends on health care professionals to report adverse incidents related to medical devices. Any serious and/or unexpected adverse incident related to medical devices should be reported to Health Canada at the following address:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Tel: The Inspectorate Hotline 1-800-267-9675
For other inquiries, please refer to contact information:
Marketed Health Products Directorate (MHPD)
Email: MHPD_DPSC@hc-sc.gc.ca
Tel: (613) 954-6522
Fax: (613) 952-7738
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
Footnotes
- Footnote 1
-
Important safety information regarding DIATHERMY THERAPY in patients with implanted leads and implanted systems with leads.
Notice To Hospitals. Health Products and Food Branch, Health Canada. April 2, 2003. - Footnote 2
-
FDA Public Health Notification: MRI-Caused Injuries in Patients with Implanted Neurological Stimulators.