Health product recall

Immulite Systems - Rapid Thyroid Stimulating Hormone (TSH) Assays

Starting date:
May 15, 2013
Posting date:
July 23, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-34725

Recalled products

  1. Immulite System - Rapid TSH Assay
  2. Immulite 2000 System - Rapid Thyroid Stimulating Hormone (TSH) Assay
  3. Immulite 1000 System - Rapid TSH Assay
  4. Immulite System - Third Generation TSH Assay
  5. Immulite 2000 System - Third Generation TSH Assay
  6. Immulite 1000 System - Third Generation TSH Assay

Reason

Siemens Healthcare Diagnostics is informing customers that a rare variant of TSH, identified in a small cluster of patients, is not detected by the siemens assays. As part of an investigation of unexplained discordant results, samples were identified in which the monoclonal antibody used in the reagent failed to detect the TSH molecule. These individuals may have a previously unrecognized, functionally normal TSH variant. The observed rate of occurrence during a 30-month time period was 0.6 x 10-7.

Affected products

A. Immulite System - Rapid TSH Assay

Lot or serial number

All lots

Model or catalog number
  • LKRT1
  • LKRT5
Companies
Manufacturer

Siemens Healthcare Diagnostics Products Ltd.

Glyn Rhonwy

Llanberis, Gwynedd

LL55 4EL

UNITED KINGDOM



B. Immulite 2000 System - Rapid Thyroid Stimulating Hormone (TSH) Assay

Lot or serial number

All lots

Model or catalog number
  • L2KRT2
  • L2KRT6
Companies
Manufacturer

Siemens Healthcare Diagnostics Products Ltd.

Glyn Rhonwy

Llanberis, Gwynedd

LL55 4EL

UNITED KINGDOM



C. Immulite 1000 System - Rapid TSH Assay

Lot or serial number

All lots

Model or catalog number
  • LKRT1
  • LKRT5
Companies
Manufacturer

Siemens Healthcare Diagnostics Products Ltd.

Glyn Rhonwy

Llanberis, Gwynedd

LL55 4EL

UNITED KINGDOM



D. Immulite System - Third Generation TSH Assay

Lot or serial number

All lots

Model or catalog number
  • LKTS1
  • LKTS5
Companies
Manufacturer

Siemens Healthcare Diagnostics Products Ltd.

Glyn Rhonwy

Llanberis, Gwynedd

LL55 4EL

UNITED KINGDOM



E. Immulite 2000 System - Third Generation TSH Assay

Lot or serial number

All lots

Model or catalog number
  • L2KTS2
  • L2KTS6
Companies
Manufacturer

Siemens Healthcare Diagnostics Products Ltd.

Glyn Rhonwy

Llanberis, Gwynedd

LL55 4EL

UNITED KINGDOM



F. Immulite 1000 System - Third Generation TSH Assay

Lot or serial number

All lots

Model or catalog number
  • LKTS1
  • LKTS5
Companies
Manufacturer

Siemens Healthcare Diagnostics Products Ltd.

Glyn Rhonwy

Llanberis, Gwynedd

LL55 4EL

UNITED KINGDOM