This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Immulite 2000 System - SHBG

Starting date:
January 31, 2017
Posting date:
February 14, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62204

Reason

Certain lots do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8 degree Celsius, as published in the instructions for use (IFU). After a two week (14 days) period a bias in patient samples may show result in quality control falling outside of established ranges. Identified lots can exhibit an average bias of 14.5% at day 21 to 16.5% at day 28 for sex hormone-binding globulin (SHBG) concentration of 5nmol/L to 170nmol/L during the open vial adjustor stability period of 30 days. The bias observed was not concentration dependent.

Affected products

Immulite 2000 System - SHBG

Lot or serial number

357, 358, 360, 361, 363

Model or catalog number

L2KSH2

Companies

Manufacturer

Siemens Healthcare Diagnostics Products Limited

Glyn Rhonwy

Llanberis

LL55 4EL

UNITED KINGDOM