Immulite 2000 System - SHBG

Starting date:
January 31, 2017
Posting date:
February 14, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62204

Reason

Certain lots do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8 degree Celsius, as published in the instructions for use (IFU). After a two week (14 days) period a bias in patient samples may show result in quality control falling outside of established ranges. Identified lots can exhibit an average bias of 14.5% at day 21 to 16.5% at day 28 for sex hormone-binding globulin (SHBG) concentration of 5nmol/L to 170nmol/L during the open vial adjustor stability period of 30 days. The bias observed was not concentration dependent.

Affected products

Immulite 2000 System - SHBG

Lot or serial number

357, 358, 360, 361, 363

Model or catalog number

L2KSH2

Companies

Manufacturer
Siemens Healthcare Diagnostics Products Limited
Glyn Rhonwy
Llanberis
LL55 4EL
UNITED KINGDOM