Immulite 2000-Instrument Class 4 & 2000 XPI Instrument (2020-05-20)
- Starting date:
- May 20, 2020
- Posting date:
- June 19, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73345
Last updated:
Affected Products
- Immulite 2000-Instrument Class 4
- Immulite 2000-2000 XPI Instrument
Reason
Siemens Healthcare Diagnostics has received customer complaints that the flexible tubing connected to the liquid waste bottle can crack during routine customer maintenance.
Affected products
A. Immulite 2000-Instrument Class 4
Lot or serial number
- C0315R
- G2115
- I3150
- J3828
- J4091
- P5687
- k4340
Model or catalog number
- 030002-13
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
62 Flanders-Bartley Road
Flanders,
07836
New Jersey
UNITED STATES
B. Immulite 2000-2000 XPI Instrument
Lot or serial number
- More than 10 numbers, contact manufacturer.
Model or catalog number
- 10373214
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
62 Flanders-Bartley Road
Flanders,
07836
New Jersey
UNITED STATES