This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Immulite 2000 and Immulite 2000 XPI (2015-04-24)

Starting date:
April 24, 2015
Posting date:
May 21, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53405

Recalled Products

  1. Immulite 2000
  2. Immulite 2000 XPI

Reason

Some systems may have been manufactured with a defective reagent carousel temperature sensor PCB or you may have had a part replacement with a defective PCB assembly.

Affected products

A. Immulite 2000

Lot or serial number

All lots.

Model or catalog number

030002-13

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

62 Flanders-Bartley Road

Flanders,

07836

New Jersey

UNITED STATES



B. Immulite 2000 XPI

Lot or serial number

All lots.

Model or catalog number
  • 030002-33
  • 10373214
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

62 Flanders-Bartley Road

Flanders,

07836

New Jersey

UNITED STATES