Health product recall

IMMAGE Immunochemistry System (2020-04-29)



Starting date:
April 29, 2020
Posting date:
May 15, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73115


Last updated:

Affecetd Products

  1. IMMAGE Immunochemistry System C3 Complement C3 Reagent
  2. IMMAGE Immunochemistry System C4 Complement C4 Reagent
  3. IMMAGE Immunochemistry System HPT Haptoglobin Reagent
  4. IMMAGE Immunochemistry System IGA Immunoglobulin A Reagent
  5. IMMAGE Immunochemistry System IGM Immunoglobulin M Reagent

Reason

Beckman Coulter has identified that IMMAGE C3, C4, HPT, IGA, and IGM reagents fail to meet IFU interference claims at high levels of lipid (triglycerides).

Affected products

A. IMMAGE Immunochemistry System C3 Complement C3 Reagent

Lot or serial number
  • All lots.
Model or catalog number
  • 446450
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



B. IMMAGE Immunochemistry System C4 Complement C4 Reagent

Lot or serial number
  • All lots. 
Model or catalog number
  • 446490
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



C. IMMAGE Immunochemistry System HPT Haptoglobin Reagent

Lot or serial number
  • All lots.
Model or catalog number
  • 446480
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



D. IMMAGE Immunochemistry System IGA Immunoglobulin A Reagent

Lot or serial number
  • All lots. 
Model or catalog number
  • 446460
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



E. IMMAGE Immunochemistry System IGM Immunoglobulin M Reagent

Lot or serial number
  • All lots.
Model or catalog number
  • 447610
Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES