Immage 800 Immunochemistry System and Reconditioned Immage (2017-02-27)

Starting date:
February 27, 2017
Posting date:
June 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63526

Affected Products

A. Immage 800 Immunochemistry System

B. Reconditioned Immage

Reason

Beckman Coulter has determined that Immage Cuvette Sectors (P/N 470706) may be missing one of two positioning pins. Cuvette sectors missing a positioning pin can become misaligned. If a cuvette sector becomes misaligned, under rare circumstances, the instrument mixer paddle may rub against the inside wall of the cuvette and create a noise.

Affected products

A. Immage 800 Immunochemistry System

Lot or serial number

All lots

Model or catalog number

A15445

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES



B. Reconditioned Immage

Lot or serial number

All lots

Model or catalog number

A36035

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

UNITED STATES