Health product recall

iLED 7 Surgical Light Systems (2019-09-07)

Starting date:
September 7, 2019
Posting date:
October 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71229


Last updated:

Affected Products

iLED 7 Surgical Light Systems

Reason

The functions (illuminance, color temperature, etc.) of the iLED7 - lighthead can be controlled via the software TruRemote running on the Control/ WallControl Panel. The control panel connects via WiFi to the access point in the light head. In some cases, the 2.4GHz-WiFi-connection is not established and control of the iLED7 from the TruRemote does not occur.

Affected products

iLED 7 Surgical Light Systems

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 4068110
  • 4068120
  • 4068210
  • 8068310
Companies
Manufacturer

Trumpf Medizin Systems GmbH T Co. Kg,

Karl-Zeiss-Str. 7-9,

Saalfeld,

07318

GERMANY