IceMan CLASSIC3™
Brand(s)
Summary
Customers are notified and provided instructions on follow up. Quarantine and return affected products.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
IceMan CLASSIC3™ |
101325 |
11-0494 |
IceMan CLASSIC3™ |
110525 |
11-0494 |
Issue
Complaints received reporting a consistent failure mode described as a broken connector on IceMan CLASSIC™ cube unit.
If the connector is broken, the unit may not function as intended, which could result in a temporary interruption of prescribed cold therapy until a replacement unit is provided. Leakage could saturate the barrier (dressings) and contact the surgical site with non-sterile water, which could in potential cold-related tissue injury or surgical site infection.