Health product recall

iBur Distal Bend Cutting Accessories

Last updated

Summary

Product
iBur Distal Bend Cutting Accessories
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

iBur Precision Match Head, Distal Bend

More than 10 numbers, contact manufacturer.

8431-107-530

iBur Diamond Round, Distal Bend

More than 10 numbers, contact manufacturer.

8431-012-040D
8431-012-020D
8431-012-030D

iBur Precision Round, Distal Bend

More than 10 numbers, contact manufacturer.

8431-009-040
8431-009-030

iBur Coarse Diamond Round, Distal Bend

More than 10 numbers, contact manufacturer.

8431-013-030DC
8431-013-050DC
8431-013-040DC

Issue

There is a potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing.

Recall Start Date: July 22, 2024

 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Stryker Instruments, A Division Of Stryker Corporation

1941 Stryker Way, Portage, Michigan, United States, 49002

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75921

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe