- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69492
Last updated:
Affected Products
- IBDoc Calprotectin
- IBDoc Calprotectin Kit
Reason
The IBDoc has an error in the design of the result upload process. Whenever a group of results is uploaded to the IBDoc Portal and one of the results in that group was already received on a separate occasion (duplicate result) then the entire group of results does not get processed further, even though the IBDoc Portal returns a success status to CalApp.