This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

iASSIST Knee System - Validation Tool

Starting date:
February 27, 2017
Posting date:
April 4, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62890

Reason

An increase of complaints was observed regarding bent or broken drive pins of the validation tool manufactured with drawing revision M to P.

Affected products

iASSIST Knee System - Validation Tool

Lot or serial number

140147

140407

141035

150120

Model or catalog number

20-8011-021-00

Companies

Manufacturer

Zimmer CAS,

75 Queen Street, Suite 3300,

Montreal

H3C 2N6

Quebec

CANADA