Hybrid Recon 4.0 and 5.0
Brand(s)
Last updated
Summary
Product
Hybrid Recon 4.0 and 5.0
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Industry
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Hybrid Recon 4.0 |
version 4.0.x* |
00859873006158 |
Hybrid Recon 5.0 |
version 5.0.0 |
VERSION 5.0 |
Issue
When reconstructing aspect/CT study, in some cases, the aspect and CT series are not correctly aligned. Normally, the hybrid recon application warns the user if the frames of reference do not match, however in some configurations this warning message may be turned off.
Recall start date: October 31, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Health products - Medical devices - Radiology
Companies
| Hermes Medical Solutions, Ab |
| Strandbergsgatan 16, Stockholm, Sweden, 11251 |
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-76498
Get notified
Receive emails about new and updated recall and safety alerts.