Hugo™ Robotic-Assisted Surgery (RAS) Tower, 120v
Brand(s)
Last updated
Summary
Product
Hugo™ Robotic-Assisted Surgery (RAS) Tower, 120v
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Hugo™ Robotic-Assisted Surgery (RAS) Tower, 120v |
C22AKC0070 |
MRASC0003 |
Issue
Reports of system error during preoperative calibration self-tests of the Hugo robotic arm cart assembly. After recalibration attempt, the robotic arm cart assembly will be prevented from entering teleoperation even if the calibration is successful. This causes procedural delays in teleoperation.
Recall start date: August 24, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Covidien LLC
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74289
Get notified
Receive notifications for new and updated recalls and alerts by category.