Health product recall

Hugo™ Robotic-Assisted Surgery (RAS) Tower, 120v

Brand(s)
Last updated

Summary

Product
Hugo™ Robotic-Assisted Surgery (RAS) Tower, 120v
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Hugo™ Robotic-Assisted Surgery (RAS) Tower, 120v

C22AKC0070

MRASC0003

Issue

Reports of system error during preoperative calibration self-tests of the Hugo robotic arm cart assembly. After recalibration attempt, the robotic arm cart assembly will be prevented from entering teleoperation even if the calibration is successful. This causes procedural delays in teleoperation.

Recall start date: August 24, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Covidien LLC

15 Hampshire Street, Mansfield, Massachusetts, United States, 02048

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74289

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe