Hugo Ras Tower, 120v
Brand(s)
Last updated
Summary
Product
Hugo Ras Tower, 120v
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Hugo Ras Tower, 120v | C22AKC0070 | MRASC0003 |
Issue
There have been 22 complaints of reported communication errors with the Hugo™ Ras System. Medtronic determined that the errors were caused by specific ports on a network switch located on the back of the Hugo™ tower. The communication errors can present in a variety of ways and may be intermittent, but the system always alerts the user of an error. In some cases, the communication errors may require a system reboot.
Recall start date: March 18, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Covidien Llc |
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75326
Get notified
Receive notifications for new and updated recalls and alerts by category.