Health product recall

Hugo RAS Tower, 120v

Brand(s)
Last updated

Summary

Product
Hugo RAS Tower, 120v
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Hugo RAS Tower, 120v

10884521826649

MRASC0003

Issue

The loss of motion has been found to occur intraoperatively due to an instrument failure, causing the system to halt teleoperation with the robotic instrument. While the instrument is disabled and during the user's removal of the instrument it is possible for the user to trigger an error code by applying excessive torque which also disables the arm cart assembly (ACA). In both scenarios (disabled instrument, and disabled ACA) halting of teleoperation is an intended response of the system to prevent further use of the robotic instrument or arm cart assembly until the issue is resolved, or the portion of the system with issue is replaced.

Recall Start Date: April 29, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Covidien Llc

15 Hampshire Street, Mansfield, Massachusetts, United States, 02048

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77429

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe