Hugo Ras Tower, 120v
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Hugo Ras Tower, 120v |
C22AKC0070 |
MRASC0003 |
Issue
There have been 22 complaints of reported communication errors with the Hugo™ Ras System. Medtronic determined that the errors were caused by specific ports on a network switch located on the back of the Hugo™ tower. The communication errors can present in a variety of ways and may be intermittent, but the system always alerts the user of an error. In some cases, the communication errors may require a system reboot.
Recall start date: March 18, 2024
Additional information
Details
| Covidien Llc |
| 15 Hampshire Street, Mansfield, Massachusetts, United States, 02048 |